Water Cooling Unit

can be installated inhouse or outside (not in direct sun exposition) providing cooling water for up to 2 analyzers, extraction machines etc. Closed water cooling system with tank and electronically controlled cooling water temperature with digital …..

Article number : 20-1144 Categories: ,

Description

Can be installed inhouse or outside (covered and not in direct sun exposition) providing cooling water for up to 2 analyzers, extraction machines etc. closed water cooling system with tank and electronically controlled cooling water temperature with digital display. In case of installation outside at temperatures below 0°C, the tank has to be filled with an anti-freezing water-/glyclol mix. The cooling capacity of the unit is sufficient for the operation of up to two Analysators at the same time. A suitable tubing installation has to be provided locally. Layout details are available on request.

Technical specifications:

  • Water tank capacity: 26 l
  • Refrigeration capacity at an ambient temperature of 32°C = 6,4kW at a water temperature of 15°C
  • Admissible ambient temperature: -20 – +42°C
  • Temperature range of circulation: 13,5°C – 25°C
  • Air capacity: 4500 m³/h
  • Flow rate of circulation pump: 2 m³/h
  • Flow height of circulation pump: 3,1 bar

To prevent corrosion and formation of biological matter it is recommended to add antifreeze (2xZ599.910) to the water filling.

Additional information

Electrical data

Dimensions approx.
Weight approx.